Global Group
Starna Potassium Dichromate Solutions
. Ultraviolet and visible regions of the spectrum (from 235 nm to 430nm) of spectrophotometers with spectral bandwidths of 2 nm or less
. The most widely used reference material for absorbance accuracy and linearity


기본구성
|
Nominal Absorbance Values at Certified Wavelengths |
||||||
|
Concentration |
235 nm |
257 nm |
313 nm |
350 nm |
430 nm |
Catalogue Number |
|
20 mg/l |
0.246 |
0.286 |
0.096 |
0.214 |
|
RM-02 |
|
40 mg/ |
0.493 |
0.574 |
0.193 |
0.428 |
|
RM-04 |
|
60 mg/l |
0.743 |
0.865 |
0.290 |
0.643 |
|
RM-06 |
|
80 mg/l |
0.994 |
1.157 |
0.387 |
0.858 |
|
RM-08 |
|
100 mg/l |
1.247 |
1.452 |
0.484 |
1.074 |
|
RM-10 |
|
120 mg/l |
1.501 |
1.749 |
0.582 |
1.290 |
|
RM-12 |
|
140 mg/l |
1.758 |
2.050 |
0.679 |
1.506 |
|
RM-14 |
|
160 mg/l |
2.016 |
2.351 |
0.777 |
1.723 |
|
RM-16 |
|
180 mg/l |
2.275 |
2.655 |
0.875 |
1.940 |
|
RM-18 |
|
200 mg/l |
2.536 |
2.960 |
0.974 |
2.157 |
|
RM-20 |
|
220 mg/l |
2.790 |
3.256 |
1.071 |
2.373 |
|
RM-22 |
|
240 mg/l |
3.043 |
3.552 |
1.168 |
2.589 |
|
RM-24 |
|
600 mg/l |
|
|
|
|
0.954 |
RM-60 |
Catalogue Identity: RM-nn where nn indicates the concentration of the solution.
For example, RM-02 is 20 mg/l, RM-04 is 40 mg/l, and so onup to RM-60, which is 600 mg/l. References can be orderedseparately or in any combination simply by combining thesenumbers: for example RM-0204060810 would be a setcomprising 20,40,60,80 and 100 mg/l references.


장비 관련 상세 문의를 원하시면 -> Click 또는 02-2659-9948 / kri@k-ri.co.kr