Global Group

Best your Partner

제품 상세보기

제품 상세보기

장비명
Starna Rare Earth Oxide Solution Cells
Wavelengths
200 – 870 nm

Starna Rare Earth Oxide Solution Cells

 

.  Covers wavelengths from the Deep UV to the visible

.  Measured at bandwidths of 0.10, 0.25, 0.50, 1.00, 1.50, 2.00, 3.00, 4.00 and 5.00 nm

  *Only the certified values should be used for instrument qualification

 

 

Far UV Reference (Cerium oxide in sulfuric acid) (200 – 270 nm)

.  Nominal certified peak wavelengths (nm): 201, 212, 223, 240, 253

 

 

 

Holmium Oxide in perchloric acid (240 -650 nm)

.  Nominal certified peak wavelength values (nm) are: 241, 250, 278, 288, 334, 346, 361, 385, 417, 451, 468, 485, 537, 641

 

 

Didymium Oxide in perchloric acid (290 – 870 nm)

.  Nominal certified peak wavelength values (nm) are: 298, 329, 354, 444, 469, 482, 512, 522, 575, 732, 740, 794, 799, 864

 

 

Combined Holmium/Didymium Reference (240 -795 nm)

.  The nominal peak wavelength values (nm) are: 241, 287, 361, 416, 444, 482, 485, 522, 537, 575, 641, 732, 740 and 794

 

 

Samarium Oxide in perchloric acid (230 – 500 nm)

.  Nominal certified peak wavelength values (nm) are: 235, 279, 290, 305, 318, 332, 344, 362, 374, 391, 401, 415, 464, 479

 

 

기본구성

 

 

CATALOGUE NUMBER

Cerium liquid cell

RM-RE

Holmium oxide liquid cell

RM-HL

Didymium oxide liquid cell

RM-DL

Holmium/Didymium liquid cell

RM-HDL

Samarium Oxide liquid cell

RM-SL

 

 

 

 

 

 

 

 

 

 

장비 관련 상세 문의를 원하시면 -> Click 또는 02-2659-9948 /  kri@k-ri.co.kr