Global Group
Basic Qualification Kits
. These kits have been designed to qualify UV-visible spectrophotometers over the wavelength and absorbance ranges commonly used in pharmaceutical analysis
기본구성
. Basic Kit for EP2.2.25 compliance
. Catalogue number: RM-EP20 [Kit Identity: RM-020610HLKISISNTX]
|
REFERENCE |
TO QUALIFY |
WAVELENGTH RANGE |
|
Holmium cell |
Wavelength accuracy |
240 - 650 nm |
|
Potassium Dichromate 20, 60 & 100 mg/l & blank cell |
Absorbance accuracy and linearity, 0-1 A |
235 – 350 nm |
|
Sodium Iodide |
Stray light, 220 nm |
220 -370 nm |
|
Potassium Iodide |
Stray light, 250 nm |
|
|
Sodium Nitrite |
Stray light, 340, 370 nm |
|
|
Toluene in Hexane |
Resolution (bandwidth), 265-270nm |
265-270 nm |
. Basic Kit for USP Chapter <857> compliance
. Catalogue number: RM-USP20 [Kit Identity: RM-020610HLACKISNTX]
|
REFERENCE |
TO QUALIFY |
WAVELENGTH RANGE |
|
Holmium cell |
Wavelength accuracy |
240 - 650 nm |
|
Potassium Dichromate 20, 60 & 100 mg/l & blank cell |
Absorbance accuracy and linearity, 0-1 A |
235 – 350 nm |
|
Potassium dichromate |
Absorbance accuracy, 1 A |
430 nm |
|
Potassium Iodide |
Stray light, 220 nm |
220 -370 nm |
|
Acetone |
Stray light, 300 nm |
|
|
Sodium Nitrite |
Stray light, 340 nm |
|
|
Toluene in Hexane |
Resolution (bandwidth), 265-270nm |
265-270 nm |
. Basic Combination kit for EP and USP compliance
. Catalogue number: RM-USPEP20 [Kit Identity: RM-020610HLACKISISNTX]
|
REFERENCE |
TO QUALIFY |
WAVELENGTH RANGE |
|
Holmium cell |
Wavelength accuracy |
240 - 650 nm |
|
Potassium Dichromate 20, 60 & 100 mg/l & blank cell |
Absorbance accuracy and linearity, 0-1 A |
235 – 350 nm |
|
Potassium dichromate |
Absorbance accuracy, 1 A (USP) |
430 nm |
|
Sodium Iodide |
Stray light, 220 nm (EP) |
220 -370 nm |
|
Potassium Iodide |
Stray light, 220 nm (USP), 250nm (EP) |
|
|
Acetone |
Stray light, 300 nm (USP) |
|
|
Sodium Nitrite |
Stray light, 340 nm (EP & USP), 370 nm (EP) |
|
|
Toluene in Hexane |
Resolution (bandwidth), 265-270nm |
265-270 nm |
. Far UV extension kit – EP and USP compliance
. Catalogue number: RM-06A12A24AREKC
|
REFERENCE |
TO QUALIFY |
WAVELENGTH RANGE |
|
Cerium cell |
Wavelength accuracy |
200 -270 nm |
|
Nicotinic acid 6, 12 & 24 mg/l |
Absorbance accuracy and linearity, 0-1 A |
210 - 270 nm |
|
Potassium Chloride |
Stray light, 200 nm |
200 nm |
장비 관련 상세 문의를 원하시면 -> Click 또는 02-2659-9948 / kri@k-ri.co.kr